Skip to main navigation Skip to search Skip to main content

Water bath sonicator integrated with pid-based temperature controller for flavonoid extraction

Research output: Contribution to journalArticlepeer-review

Abstract

In this research, water bath sonicator was fabricated to extract bioactive compound of plants material using sound energy (ultrasonic waves) and heater. The bioactive compound, flavonoid, has high sensitivity to temperature and extraction time and previous research stated best treatment with combination of 45C and 20 minutes. Therefore, fabricated water bath sonicator was equipped with proportional integral derivative (PID) based temperature controller and timer. Based on a calculation using the Ziegler-Nichols tuning method, Kp, Ki, Kd parameters are 16.59, 0.0279, and 2463.6, respectively. The experimental result shows that the PID controller can perform as design specification with overshoot 1.39%, error steady-state 0.688% and settling time 37.2 minutes. Furthermore, it was proven that the PID controller has contribution to extract more flavonoid.

Original languageEnglish
Article number14830
Pages (from-to)970-976
Number of pages7
JournalTelkomnika (Telecommunication Computing Electronics and Control)
Volume18
Issue number2
DOIs
Publication statusPublished - 1 Apr 2020

Keywords

  • Flavonoid extraction
  • PID controller
  • Temperature
  • Timer
  • Water bath sonicator

Fingerprint

Dive into the research topics of 'Water bath sonicator integrated with pid-based temperature controller for flavonoid extraction'. Together they form a unique fingerprint.